Features:
1. LCD display
2. Double control modes of bed temperature and skin temperature
3. Double-layer thermostatic cover, automatic air curtain device
4. Humidity control function
5. Air & Oxygen blender
6. Oxygen supply system
7. Baby cot is infinitely adjustable
8. 24 kinds of sound and light alarm, comprehensive protection of baby safety
Technical Parameters:
|
General Parameters |
Item |
Specifications |
|
Power Supply |
AC220V, 50/60Hz(Standard) 1320VA; 110V 60Hz(Optional) |
|
|
Size(L*W*H) |
950*645*(1250-11450) mm |
|
|
Maximum Heater Output Power(220VAC) |
Air chamber heater: 330W; Water tank heater: 155W |
|
|
Auxiliary Network Power Output |
AC220V/50Hz, both maximum permissible currents are 220V 1.5A |
|
|
Heater Indication |
0~100%, 10% increasing amount |
|
|
Temperature Control Mode |
Bed temperature mode; The skin temperature mode |
|
|
Bed Temperature Control Range |
25.0~37.0℃; 37.1~39.0℃(Span mode) |
|
|
Skin Temperature Control Range |
34.0~37.0℃; 37.1~38.0℃(Span mode) |
|
|
Skin Temperature Sensor Display Range |
5.0~65.0℃ |
|
|
Heating Time |
≤50min |
|
|
Temperature Variability |
≤0.5℃ |
|
|
Temperature Uniformity (Mattress in horizontal position) |
≤0.8℃ |
|
|
Temperature Uniformity (Mattress in tilted position) |
≤1.0℃ |
|
|
Difference Between Average Incubator Temperature And Control Temperature |
≤1.5℃ |
|
|
Skin Temperature Sensor Accuracy |
±0.3℃ |
|
|
Humidity Control |
The Running Time Of The Tank After It Is Filled With Distilled Water |
Humidity below 70%RH, at least 24h |
|
Water Tank Capacity |
1500ml |
|
|
Humidity Display Range |
0%RH~99%RH |
|
|
Humidity Control Range |
CAH%RH~90%RH(1% per adjustment) |
|
|
Humidity Control Accuracy |
±10% |
|
|
Oxygen Concentration Control |
Oxygen Concentration Display Range |
0%~99% |
|
Oxygen Concentration Display Resolution |
1% |
|
|
Oxygen Concentration Control Accuracy |
±5% oxygen volume concentration |
|
|
Oxygen Concentration Setting Range |
21%~60% (1% per adjustment) |
|
|
Oxygen Sensor Service Life |
The maximum use time is 10000h at 100% oxygen concentration |
|
|
Transport Storage |
Temperature |
-20℃~+55℃ |
|
Atmospheric Pressure Range |
500hPa~1060hPa |
|
|
Relative Humidity |
≤93%RH |
|
|
Packing Size/ Gross Weight |
1150*800*1600mm 190kg |
|
|
The Work Environment |
Relative Humidity |
30%RH~75%RH |
|
Environment Temperature |
+20℃~+30℃ |
|
|
Pressure Range |
700hPa~1060hPa |
|
|
Service Altitude |
≤2000m |
|
|
Overvoltage Category |
Ⅱ |
|
|
Pollution Level |
2 |
|
|
Ambient Air Velocity |
≤ 0.3m/s |
|
|
Other Indicators |
Ambient Noise In The Infant Compartment Of The Incubator |
No more than 55dB(A) at stable temperature |
|
Carbon Dioxide Concentration In The Infant Compartment |
A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL /min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0 5% after stabilization |
|
|
The Velocity Of Air Flow Over The Baby Mattress |
≤0.35m/s |
|
|
General Alarm |
Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm |
|
|
System Alarm |
Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure |
|
Configuration:
|
Model |
BK-3201(I) |
BK-3201(II) |
|
Temp. Control |
Bed/Skin temperature |
Bed/Skin temperature |
|
Display |
LCD |
LCD |
|
Bed Temp. Control Range |
25℃-39℃ |
25℃-39℃ |
|
Skin Temp. Control Range |
34℃-38℃ |
34℃-38℃ |
|
Humidity Control Range |
CAH%~99%RH |
CAH%~99%RH |
|
Oxygen Control Range |
21%-99% |
/ |
|
Cabinet |
Yes |
Yes |
Share