Skip to product information
1 of 1

Express Bridge

MicroPette Mechanical Pipette 0.1-100μl

MicroPette Mechanical Pipette 0.1-100μl

Regular price 1 som
Regular price Sale price 1 som
Sale Sold out

All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22℃.

Features for Micropette:

* Light weight, ergonomic, low force design

* Digital display clearly reads volume setting

* The pipettes cover volume range of 0.1μl to 10ml

* Easy to calibrate and maintain 

* The lower part is autoclavable

* Design helps avoid repetitive strain injuries

* Manufactured from innovative materials

Features for Multi-channel Micropette:

* 8 and 12 channel pipettes are available for standard 96-well plate 

* Dispensing head rotates ensure optimum pipetting convenience 

* Individual piston/tip cone assemblies allow easy repair and maintenance 

* Compound material tip cone design allows visual seal vertification 

  • Universal pipette tips can be used 
  • odels:

    Single-channel Adjustable Volume Pipettes

    Volume Range

    Increment

    Test Volume

    (μl)

    Maximume Permissible Systematic error

       (Inaccuracy)

    Maximume Permissible

    Random error

       (Inaccuracy)




    %

    μl

    %

    μl

    0.1-2.5μl

    0.05μl

    2.5

    2.50%

    0.0625

    2.00%

    0.05

    1.25

    3.00%

    0.0375

    3.00%

    0.0375

    0.25

    12.00%

    0.03

    6.00%

    0.015

    0.5-10μl

    0.1μl

    10

    1.00%

    0.1

    0.80%

    0.08

    5

    1.50%

    0.075

    1.50%

    0.075

    1

    2.50%

    0.025

    1.50%

    0.015

    2-20μl

    0.5μl

    20

    0.90%

    0.18

    0.40%

    0.08

    10

    1.20%

    0.12

    1.00%

    0.1

    2

    3.00%

    0.06

    2.00%

    0.04

    5-50μl

    0.5μl

    50

    0.60%

    0.3

    0.30%

    0.15

    25

    0.90%

    0.225

    0.60%

    0.15

    5

    2.00%

    0.1

    2.00%

    0.1

    10-100μl

    1μl

    100

    0.80%

    0.8

    0.15%

    0.15

    50

    1.00%

    0.5

    0.40%

    0.2

    10

    3.00%

    0.3

    1.50%

    0.15

    20-200μl

    1μl

    200

    0.60%

    1.2

    0.15%

    0.3

    100

    0.80%

    0.8

    0.30%

    0.3

    20

    3.00%

    0.6

    1.00%

    0.2

    50-200μl

    1μl

    200

    0.60%

    1.2

    0.15%

    0.3

    100

    0.80%

    0.8

    0.30%

    0.3

    50

    1.00%

    0.5

    0.40%

    0.2

    100-1000μl

    5μl

    1000

    0.60%

    6

    0.20%

    2

    500

    0.70%

    3.5

    0.25%

    1.25

    100

    2.00%

    2

    0.70%

    0.7

    200-1000μl

    5μl

    1000

    0.60%

    6

    0.20%

    2

    500

    0.70%

    3.5

    0.25%

    1.25

    200

    0.90%

    1.8

    0.30%

    0.6

    1000-5000μl

    50μl

    5000

    0.50%

    25

    0.15%

    7.5

    2500

    0.60%

    15

    0.30%

    7.5

    1000

    0.70%

    7

    0.30%

    3

    2-10ml

    0.1ml

    10ml

    0.60%

    60

    0.20%

    20

    5ml

    1.20%

    60

    0.30%

    15

    2ml

    3.00%

    60

    0.60%

    12


    8-channel Adjustable Volume Pipettes

    Volume Range

    Increment

    Test Volume

    (μl)

    Maximume Permissible Systematic error

      (Inaccuracy)

    Maximume Permissible Random error

    (Inaccuracy)




    %

    μl

    %

    μl

    0.5-10μl

    0.1μl

    10

    1.50%

    0.15

    1.50%

    0.15

    5

    2.50%

    0.125

    2.50%

    0.125

    1

    4.00%

    0.04

    4.00%

    0.04

    5-50μl

    0.5μl

    50

    1.00%

    0.5

    0.50%

    0.25

    25

    1.50%

    0.375

    1.00%

    0.25

    5

    3.00%

    0.15

    2.00%

    0.1

    50-300μl

    5μl

    300

    0.70%

    2.1

    0.25%

    0.75

    150

    1.00%

    1.5

    0.50%

    0.75

    50

    1.50%

    0.75

    0.80%

    0.4

    12-channel Adjustable Volume Pipettes

    Volume Range

    Increment

    Test Volume

    (μl)

    Maximume Permissible Systematic error

       (Inaccuracy)

    Maximume Permissible Random error

    (Inaccuracy)




    %

    μl

    %

    μl

    0.5-10μl

    0.1μl

    10

    1.50%

    0.15

    1.50%

    0.15

    5

    2.50%

    0.125

    2.50%

    0.125

    1

    4.00%

    0.04

    4.00%

    0.04

    5-50μl

    0.5μl

    50

    1.00%

    0.5

    0.50%

    0.25

    25

    1.50%

    0.375

    1.00%

    0.25

    5

    3.00%

    0.15

    2.00%

    0.1

    50-300μl

    5μl

    300

    0.70%

    2.1

    0.25%

    0.75

    150

    1.00%

    1.5

    0.50%

    0.75

    50

    1.50%

    0.75

    0.80%

    0.4

    Fixed Volume Pipettes

    Volume Range

    Increment

    Test Volume

    (μl)

    Maximume Permissible Systematic error

       (Inaccuracy)

    Maximume Permissible Random error

    (Inaccuracy)




    %

    μl

    %

    μl

    5μl

    -

    5μl

    1.3%

    0.065

    1.2%

    0.06

    10μl

    -

    10μl

    0.8%

    0.08

    0.8%

    0.08

    20μl

    -

    20μl

    0.6%

    0.12

    0.5%

    0.1

    25μl

    -

    25μl

    0.5%

    0.125

    0.3%

    0.075

    50μl

    -

    50μl

    0.5%

    0.25

    0.3%

    0.15

    100μl

    -

    100μl

    0.5%

    0.5

    0.3%

    0.3

    200μl

    -

    200μl

    0.4%

    0.8

    0.2%

    0.4

    250μl

    --

    250μl

    0.4%

    1.0

    0.2%

    0.5

    500μl

    -

    500μl

    0.3%

    1.5

    0.2%

    1.0

    1000μl

    -

    1000μl

    0.3%

    3.0

    0.2%

    2.0

    2000μl

    -

    2000μl

    0.3%

    6.0

    0.15%

    3.0

    5000μl

    -

    5000μl

    0.3%

    15

    0.15%

    7.5

 

View full details